  <regulation prefix="Regulation" num=".01" heading="Definitions.">
    <paragraph num="A.">
      <text>In this chapter, the following terms have the meanings indicated.</text>
    </paragraph>
    <paragraph num="B.">
      <text>Terms Defined.</text>
      <paragraph num="(1)">
        <text>Critical Items.</text>
        <paragraph num="(a)">
          <text>“Critical items” means the instruments or objects that come in direct contact with the bloodstream or other normally sterile areas of the body.</text>
        </paragraph>
        <paragraph num="(b)">
          <text>“Critical items” includes tweezers and needles.</text>
        </paragraph>
      </paragraph>
      <paragraph num="(2)">
        <text>“Transfer forceps” means a medical tool that is used for grasping or holding things.</text>
      </paragraph>
    </paragraph>
  </regulation>
  <regulation prefix="Regulation" num=".02" heading="Generally.">
    <paragraph num="A.">
      <text>Each electrologist office and institution teaching the clinical practice of electrology shall maintain and use sterilization equipment adequate in size to accommodate critical items, transfer forceps, and soaking receptacles.</text>
    </paragraph>
    <paragraph num="B.">
      <text>Except as provided in §C of this regulation, critical items, transfer forceps, and soaking receptacles that are dropped on the floor or otherwise contaminated may not be used until sterilized.</text>
    </paragraph>
    <paragraph num="C.">
      <text>All needles used in an electrologist office and institution teaching the clinical practice of electrology shall be purchased as pre-sterilized and disposable for one-time use only. The sterilization procedure outlined in Regulation .04 of this chapter is not applicable to needles. If a needle is dropped on the floor or otherwise contaminated, the needle shall be properly disposed of and may not be used.</text>
    </paragraph>
  </regulation>
  <regulation prefix="Regulation" num=".03" heading="Cleansing Procedures.">
    <text>Prior to sterilization as required in Regulation .03 of this chapter, tweezers and needle caps shall be:</text>
    <paragraph num="A.">
      <text>Mechanically precleaned as follows:</text>
      <paragraph num="(1)">
        <text>Cleansed with a clean cotton ball or swab moistened with a low-residue detergent and cool water;</text>
      </paragraph>
      <paragraph num="(2)">
        <text>Accumulated in a holding container by submersion in a low-residue detergent and cool water; and</text>
      </paragraph>
      <paragraph num="(3)">
        <text>Then, thoroughly rinsed with warm water and drained;</text>
      </paragraph>
    </paragraph>
    <paragraph num="B.">
      <text>Then, cleaned by soaking in a protein-dissolving detergent enzyme cleaner or in an ultrasound cleaning unit according to manufacturer's instructions; and</text>
    </paragraph>
    <paragraph num="C.">
      <text>Rinsed and dried.</text>
    </paragraph>
  </regulation>
  <regulation prefix="Regulation" num=".04" heading="Sterilization.">
    <paragraph num="A.">
      <text>Except as provided in Regulation .02C of this chapter, critical items, transfer forceps, and soaking receptacles shall be placed on a steel tray or rack in the center of an autoclave or heat sterilizer, separating instruments and keeping them away from the walls of the unit, and shall be sterilized according to manufacturer specifications or by:</text>
      <paragraph num="(1)">
        <text>Autoclave at the temperatures listed for the following minimum time periods after completing a heat-up period:</text>
        <paragraph num="(a)">
          <text>15—20 minutes at 121°C (250°F), 15 psi (pounds per square inch) for unpackaged instruments; and</text>
        </paragraph>
        <paragraph num="(b)">
          <text>30 minutes at 121°C (250°F), 15 psi (pounds per square inch) for packaged instruments; or</text>
        </paragraph>
      </paragraph>
      <paragraph num="(2)">
        <text>Dry heat, at the temperatures listed for the following minimum time periods after completing a heat-up period:</text>
        <paragraph num="(a)">
          <text>170°C (340°F) for 1 hour; and</text>
        </paragraph>
        <paragraph num="(b)">
          <text>160°C (320°F) for 2 hours.</text>
        </paragraph>
      </paragraph>
    </paragraph>
    <paragraph num="B.">
      <text>Sterile critical items, transfer forceps, and soaking receptacles shall be stored in sterile covered containers until used.</text>
    </paragraph>
    <paragraph num="C.">
      <text>Transfer forceps and holding containers shall be cleaned and sterilized daily or more frequently if visibly contaminated.</text>
    </paragraph>
    <paragraph num="D.">
      <text>Unused instruments in containers and containers that have been opened shall be resterilized after a 24-hour period.</text>
    </paragraph>
    <paragraph num="E.">
      <text>Dry heat ovens and autoclaves shall be of a type registered with the federal Food and Drug Administration (FDA) and shall be cleaned, used, and maintained according to manufacturers' instructions.</text>
    </paragraph>
    <paragraph num="F.">
      <text>Sterilizers shall have visible temperature and time indicators.</text>
    </paragraph>
    <paragraph num="G.">
      <text>Chemical indicators using color change materials shall be used on or in each package to assure that each package or container has been processed through the sterilization cycle.</text>
    </paragraph>
    <paragraph num="H.">
      <text>Biological indicators shall be used no less than once a month according to manufacturers’ instructions to assure that the sterilizer is functioning properly.</text>
    </paragraph>
  </regulation>
  <annotations>
    <annotation type="Authority">Health Occupations Article, §§8-205 and 8-6B-01—8-6B-29, Annotated Code of Maryland</annotation>
    <annotation type="History" subtype="Administrative History">Effective date: January 16, 1984 (11:1 Md. R. 45)</annotation>
    <annotation type="History" subtype="Administrative History">Regulation .01 amended effective August 5, 1991 (18:15 Md. R. 1727)</annotation>
    <annotation type="History" subtype="Administrative History">Regulation .01, Requirements for Instructor's License, repealed and new Regulations .01—.03 adopted under new chapter, Sterilization Procedures, effective June 20, 2005 (32:12 Md. R. 1047)</annotation>
    <annotation type="History" subtype="Administrative History">Chapter revised effective August 5, 2024 (51:15 Md. R. 710)</annotation>
    <annotation type="History" subtype="Administrative History">Regulation.04A amended effective September 15, 2025 (52:18 Md. R. 923)</annotation>