<?xml version='1.0' encoding='utf-8'?>
<regulation prefix="Regulation" num=".06" citation="COMAR 30.09.10.06" title_num="30" subtitle_num="09" chapter_num="10" heading="Standard Ambulance Suction Units.">
  <text>Each ambulance shall be equipped with one:</text>
  <paragraph num="A.">
    <text>Onboard suction unit which is:</text>
    <paragraph num="(1)">
      <text>Powered by a reliable source from:</text>
      <paragraph num="(a)">
        <text>The engine manifold, or</text>
      </paragraph>
      <paragraph num="(b)">
        <text>An electrically powered vacuum pump;</text>
      </paragraph>
    </paragraph>
    <paragraph num="(2)">
      <text>Capable of developing a:</text>
      <paragraph num="(a)">
        <text>Vacuum pressure equivalent to at least 300 mm Hg within 4 seconds of clamping, and</text>
      </paragraph>
      <paragraph num="(b)">
        <text>Free air flow of at least 20 liters/minute at the end of the collection tube;</text>
      </paragraph>
    </paragraph>
    <paragraph num="(3)">
      <text>Adjustable in suction force for use in children and intubated patients;</text>
    </paragraph>
    <paragraph num="(4)">
      <text>Equipped with wide bore tubing;</text>
    </paragraph>
    <paragraph num="(5)">
      <text>Equipped with one rigid pharyngeal suction tip; and</text>
    </paragraph>
    <paragraph num="(6)">
      <text>Equipped with two suction catheters with integral thumb controls in assorted sizes appropriate for the patients being transported including at least one size from among:</text>
      <paragraph num="(a)">
        <text>6, 8, 10, or 12 French catheter, and</text>
      </paragraph>
      <paragraph num="(b)">
        <text>14, 16, or 18 French catheter; and</text>
      </paragraph>
    </paragraph>
  </paragraph>
  <paragraph num="B.">
    <text>Portable suction unit which is:</text>
    <paragraph num="(1)">
      <text>Capable of operating continuously under suction for at least 20 minutes while independent of alternating current;</text>
    </paragraph>
    <paragraph num="(2)">
      <text>Capable of developing:</text>
      <paragraph num="(a)">
        <text>A vacuum pressure equivalent to 300 mm Hg within 4 seconds of clamping, and</text>
      </paragraph>
      <paragraph num="(b)">
        <text>A free air flow of at least 20 liters/minute at the end of the suction tube;</text>
      </paragraph>
    </paragraph>
    <paragraph num="(3)">
      <text>Equipped with wide bore tubing;</text>
    </paragraph>
    <paragraph num="(4)">
      <text>Equipped with one rigid pharyngeal suction tip; and</text>
    </paragraph>
    <paragraph num="(5)">
      <text>Equipped with two suction catheters with integral thumb controls in sizes appropriate for the patients being transported including one size from among:</text>
      <paragraph num="(a)">
        <text>6, 8, 10, or 12 French catheter, and</text>
      </paragraph>
      <paragraph num="(b)">
        <text>14, 16, or 18 French catheter.</text>
      </paragraph>
    </paragraph>
  </paragraph>
</regulation>